BONDCHECK Specification |
||||||
Display |
Type |
5.7” (145mm), 18 bit Colour, daylight readable. |
||||
Viewable Area |
115.2mm (Horizontal) x 86.4mm (Vertical). Resolution 640 x 480 pixels |
|||||
Colour Schemes |
User configurable Dark, Bright and Black & White. |
|||||
Configurable Screen |
Full Screen, Single, Dual Pane with variable size and location and function e.g. XY, Timebase. |
|||||
Display Modes |
Pitch-Catch, Resonance & MIA: Spot and Sweep modes. RF Timebase for Pitch-Catch only. |
|||||
Graticules |
None, Grid (4 sizes 5, 10, 15 and 20% FSH), Polar (4 sizes 5, 10, 15 and 20% FSH) |
|||||
Offset |
Spot Position: Y -50 to +50, X -65 to +65% |
|||||
Flip |
Manual or automatic screen orientation change to enable left or right handed use. |
|||||
Transmit |
Operating Mode |
Pitch-Catch, Resonance and MIA. |
||||
Output Frequency Range |
Pitch Catch: 5kHz to 50kHz. |
Resonance: 50kHz to 400kHz |
MIA: 2kHz to 10kHz |
|||
Output Voltage |
Pitch-Catch tone burst: 10 ranges: 1,3,6,8,10,12,18,24,30,36V pk-pk. |
|||||
Pitch-Catch sweep: 3 ranges: 12, 24, 36V pk-pk |
||||||
Resonance: 3 ranges: 12, 24, 36V pk-pk |
||||||
MIA: 3 ranges: 12, 24, 36Vpk-pk (high voltage drive in probe) |
||||||
Minimum Output drive load impedance |
300 Ohms |
|||||
Waveform Type: Pitch-Catch/MIA |
Tone burst with rectangular or hanning window with chirp. |
|||||
Transmit waveform points maximum: 8192 |
||||||
Waveform duration: Maximum 3.2ms / 2.5ms |
Waveform output DAC clock rate: 2.5MHz fixed |
|||||
Frequency Sweep: Frequency range 5kHz to 50kHz / 2kHz to 10kHz |
||||||
Waveform Type: Resonance |
Fixed or swept waveform |
|||||
Frequency range 5kHz to 400 kHz |
||||||
Receive |
Pitch-Catch / MIA Tone Burst |
Sample rate: 440kHz / 100kHz |
Maximum PRF: 14Hz |
|||
Sample Bit depth: 12 bit |
Gain range: 0 to 60dB |
|||||
Receive bandwidth: 5kHz to 100kHz -6dB points |
||||||
Input voltage saturation: ±400mV |
||||||
Time base range: 100μs to 2ms/ 22ms |
Time base delay: 0μs to 1ms |
|||||
Cross Talk: >40dB isolation |
Amplitude/phase extraction cursor: position resolution <5μs /10μs |
|||||
Resonance & Pitch-Catch Sweep |
Dynamic Range: >150dB |
|||||
Bit depth: 24 bit |
||||||
Gain Range: -30 to 60dB. |
||||||
Receive bandwidth: DC to 20MHz |
||||||
Amplitude/phase extraction by QAM demodulation |
||||||
Filtering |
Fixed Hardware High-pass filter for Pitch-Catch to reduce surface scanning noise. |
|||||
Fixed Hardware Low-pass filters 100kHz for Pitch-Catch for optimum amplifier SNR. |
||||||
Configurable Software High-pass and Low-pass filters for all modes |
||||||
Software |
Acquisition Gate in RF Mode |
Adjustable gate start, width and threshold. |
||||
Alarm Gate in Y-T Mode |
Multiple box, circle and sector alarm zones. |
|||||
Calibration Mode |
Performs frequency sweep of bond and dis-bonded areas. |
|||||
Bond/Dis-bond Alarm |
Status on screen and probe LED. |
|||||
Physical |
Weight |
1.2 kg, 2.7 lbs. |
||||
Size (w x h x d) |
237.5mm x 144mm x 52mm / 9.4” x 5.7” x 2.1” |
|||||
Material/Housing |
Aluminium alloy Mg Si 0.5 powder-coated |
|||||
Operating/ Storage Temperature |
Operating 0 to +40 °C |
Storage for up to 12 months -20 to +60 °C. Nominal +20 °C |
||||
IP Rating |
54 |
|||||
Warranty |
2 year Manufacturer’s Warranty |
Covers all components of the BondCheck, excludes customer damage or misuse. (probes not included) |
Bond Check – Bond Testing
Rp98.000.000
Easy to use menus and icon system.
The BondCheck menu system is simple and fast to navigate. It has the ability to add individually selectable soft key menu items to the sidebar for rapid function access and a “quick-setting menu” for easy set-up, review and adjustment.
- Support Pitch-Catch, Resonance adn MIA Bond-Testing Modes
- Pitch and Catch Dry Coupled Bond Testing mode for rapid detection of defects in laminate, bonded and sandwich structures.
- Automatic Test Frequency Optimisation.
- Sweep, RF, T/T, Encoded and Phase/Frequency Plots.
If you have questions feel free contact us :
sales@dfn-teknologi.com or WhatsApp for fast response